Télécharger Imprimer la page

Microlife MAM BP A3 Easy Manuel D'instructions page 6

Publicité

Les langues disponibles

Les langues disponibles

8. When the correct pressure is reached, the pumping stops and
the pressure falls gradually. If the required pressure was not
reached, the device will automatically pump some more air into
the cuff.
9. During the measurement, the pulse indicator AP flashes in the
display.
10.The result, comprising the systolic AK and the diastolic AL
blood pressure and the pulse rate AM are displayed. Note also
the explanations on further display symbols in this booklet.
11.When the device has finished measuring, remove the cuff.
12.Switch off the device. (The monitor does switch off automati-
cally after approx. 1 min.).
You can stop the measurement at any time by pressing the
ON/OFF button or open the cuff (e.g. if you feel uneasy or
an unpleasant pressure sensation).
If the systolic blood pressure is known to be very high,
it can be an advantage to set the pressure individually.
Press the ON/OFF button after the monitor has been
pumped up to a level of approx. 30 mmHg (shown on the
display). Keep the button pressed until the pressure is
about 40 mmHg above the expected systolic value – then
release the button.
How do I evaluate my blood pressure
The triangle on the left-hand edge of the display AN points at the
range within which the measured blood pressure value lies. The
value is either within the optimum (green), elevated (yellow) or
high (red) range. The classification corresponds to the following
ranges defined by international guidelines (ESH, ESC, JSH). Data
in mmHg.
Range
Systolic Diastolic Recommendation
≥135
1. blood pressure
too high
2. blood pressure
130 - 134 80 - 84
elevated
3. blood pressure
<130
normal
The higher value is the one that determines the evaluation.
Example: a blood pressure value of 140/80 mmHg or a value of
130/90 mmHg indicates «blood pressure too high».
4
≥85
Seek medical
advice
Self-check
<80
Self-check
4. Appearance of the irregular heartbeat (IHB) symbol
This symbol AS indicates that an irregular heartbeat was detected.
In this case, the measured blood pressure may deviate from your
actual blood pressure values. It is recommended to repeat the
measurement.
Information for the doctor in case of repeated appearance of
the IHB symbol:
This device is an oscillometric blood pressure monitor that also
measures the pulse during blood pressure measurement and
indicates when the heart rate is irregular.
5. Data memory
This device always stores the last result at the end of the measure-
ment. To recall the reading, press and hold the ON/OFF button 1
while the device is switched off. Release the button as soon as all
display elements are shown. The stored result will then be shown,
indicated with the letter «M» BK.
6. Battery indicator and battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
AT will flash as soon as the device is switched on (partly filled
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.
Flat battery - replacement
When the batteries are flat, the battery symbol AT will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 4 on the bottom of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
Which batteries and which procedure?
Use 4 new, long-life 1.5 V, size AA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for
a prolonged period.

Publicité

loading