RADIOMETER TOSCA 500 Manuel D'utilisation page 113

Publicité

TOSCA
EXTERNAL CONNECTIONS
500
15.7 Analog outputs
PCO
analog output on pin 10:
2
0 to 5 V = 0 to 125 mmHg (40mV / mmHg) or 0 to 25.0 kPa (20 mV / 0.1 kPa)
SPO
analog output on pin 11:
2
0 to 5 V = 0 to 100% (50 mV / %SPO2)
Pulse Rate analog output on pin 12:
0 to 5 V = 0 to 250 BPM (20 mV / BPM)
Pulse Wave analog output on pin 13:
0 to 5 V auto scaled
15.7.1
Calibration of an analog recording system
The analog output signals can be used for the calibration of any analog recording system
(e.g. polysomnograph in sleep labs).The analog outputs on the System Connector are set
to full scale of 5V (+/- 15mV) during the system check.
To perform a system check:Press both
and
when no parameter setting is in
progress (see section 10.2).
The analog outputs are set to "zero" when the connected sensor is placed into the calibration
chamber (no PCO
value is displayed.)
2
15.8 Nurse call
The nurse call feature is activated if any of the following conditions are true:
– One or more alarms are active and not muted.
– The auditory alarm is not suspended.
– The auditory alarm is not inhibited.
Relais:
Shorts pin 19 to pin 18 (common) if calling
Shorts pin 37 to pin 18 if not calling
15.9 Connection to external battery
Connector type:
Redel SA. Type PAG-M02 GLAC GSG
Polarity:
WARNING
If the external battery is connected to a battery recharging device, this
device must be medical grade (double isolation).
111