EXTERNAL CONNECTIONS
Note
The following alarm philosophy is used on the Philips Monitor with the TOSCA 500 VueLink
Module:
– The alarm messages described above are "yellow" priority alarms.
– The alarm signals are indicated on the display of the Philips Monitor by an alarm message
on the center of the upper line and by the blinking of the related measured value, if
displayed.
– The alarm signals disappear on the Philips Monitor if the related auditory alarm signals on
the TOSCA 500 monitor are muted.
– Only one of the alarm messages described above is displayed at time. If more than one
alarm condition is present, only the alarm message with the highest priority is displayed.
INOP messages
Philips Monitor INOP
"TOSCA MONIT. FAULT"
"TOSCA SENSOR FAULT"
"CONNECT TOSCA SENS"
"TOSCA SENSOR OFF"
"TOSCA AMBI. LIGHT"
"TOSCA INTERFERENCE"
"TOSCA PULSE SEARCH"
"TOSCA TIME ELAPSED"
"TOSCA BATTERY LOW"
"CAL. TOSCA SENSOR"
"REMEMBRANE TOSCA S"
"TOSCA NO CAL GAS"
Note
The following INOP philosophy is used on the Philips Monitor with the VueLink Module:
– An INOP message is indicated on the left of the upper line on the Philips Monitor display
as long as the related visual message is displayed on the TOSCA 500 monitor.
– Only one of the INOP messages described above is displayed at time. If more than one
INOP condition is present, only the INOP message with the highest priority is displayed.
106
corresponding text in 7
"Monitor fault" (highest priority)
"Sensor fault"
"Connect sensor" or "Unknown sensor" or "Update
software"
"Sensor off patient"
"Too much ambient light"
"Interference"
"Searching pulse signal"
"Site time elapsed"
"Battery is low"
"Calibrate sensor" or "Place sensor in cal. chamber"
"Remembrane sensor"
"No gas flow, renew gas bottle" (lowest priority)
TOSCA
500