Column pos. 6: PR value
Heading line:
PR[bpm]<limit>,
PR[bpm]050:150,
Data line:
<value>,
087,
---,
Column pos. 7: PR alarm status
Heading line:
PRs,
Data line:
<code>,
N,
A,
S,
Column pos. 8: HPWR value
Heading line:
HPWR[mW]<HPWRref
>,
HPWR[mW]256,
Data line:
<value>,
252,
---,
Column pos. 9: Perfusion Index (PI)
Heading line:
Perf.Index[%],
Data line:
<value>,
12.3,
Column pos. 10: Sensor temperature
Heading line:
Temp[C]<set>,
Temp[C]42.0,
Data line:
<value>,
41.9,
Column pos. 11: Site timer
Heading line:
Timer,
Data line:
<timer>,
02:15,
Column pos. 12: Mode
Heading line:
Mode,
Data line:
<mode>,
C,
R,
P,
M,
Indication of the PR alarm limits "low:high" valid
for the next data lines.
Indication of "---"if no value is available.
No heading parameter
Possible codes are:
"N" no alarm
"A" alarm active
"S" alarm muted, reset
Indication of the <HPWRref> (relative heating
power reference) valid for the next data lines.
Indication of "---"if no value is available.
No heading parameter
Indication of "--.-"if no value is available.
Indication of the temperature set valid for the next
data lines.
Indication of "--.-"if no value is available.
No heading parameter
Site time left of the actual data line, format
"hh:mm"
No heading parameter
Possible modes are:
"C" calibrating
"R" ready to use
"P" pre-measurement
"M" measurement
"?"
other (connect sensor, monitor fault etc.)
EXTERNAL CONNECTIONS