EXTERNAL CONNECTIONS
Column pos. 13: General status
Heading line:
Status<CR><LF>
Status
Data line:
<code><CR><LF>
N
A
S
U
Bytes per line: heading line = 133 bytes, data line = 53 bytes
Column heading line example:
"$2006.07.31,PCO2[kPa]_01.2:12.3,PCO2s,SPO2[%]88:100,SPO2s,PR[bpm]050:150,PRs,
HPWR[mW]541,Perf.Index[%],Temp[°C]44.0,Timer,Mode,Status"
Data line example:
"11:22:33,12.3,N,100,N,245,N,999,12.3,44.3,12:34,M,A"
15.5 VueLink
As part of a Philips Patient Monitoring Systems (Philips V24/26, Philips CMS and Philips
IntelliVue, here after called "Philips Monitor"), the TOSCA 500 monitor is compatible with
the following VueLink Open Interface module:
Module type
Philips VueLink module M1032A #A05 #K6B (type B, Auxiliary-plus, with Open Interface
cable).
Ordering information:
M1032A #A05
M1032A #K6B
Connection
Connect the VueLink module to the TOSCA 500 Systems Connector with the VueLink Open
Interface cable (available from Philips) and the TOSCA VueLink adapter cable (part No 500
0100).
Operation
When the VueLink protocol is selected (see section 7.2.5), the VueLink module plugged in to
the Philips Monitor Module device will automatically recognize the TOSCA 500 monitor and
will appear as "TOSCA" on the VueLink module setup menu.
See section "VueLink" of the Philips Monitor documentation for setting and configuration of
the VueLink module.
104
VueLink Open Interface Module (type B, auxiliary-plus)
VueLink Open Interface Cable, 2m, Standard 25 pin male connector
No heading parameter
This is the last column of the heading line, ending
with "CR LF" (0x0D 0x0A) bytes.