10
a
h
h
a
a
h
h
InPro 86X0 i e
30 421 372 D
Only such maintenance work as is specifically described in this instruction manual may be
performed on the sensor.
Only original spare parts from METTLER TOLEDO may be used for replacing defective components
(see chapter 7).
Danger! Non-compliance with the prescribed maintenance instructions can endanger
personnel and the environment.
2.6.2 Installation in pressurized systems
Notice: The maximum permissible temperature and pressure specifications must not
be exceeded. The relative specifications are given on the type plate of the sensor.
2.7
Residual hazards
2.7.1 Rupture of connections
Sensor connections can become loosened through the effects of vibration.
Notice: All sensor interfaces including process adapter, and the interfaces between optical
window and sensor head, need to be checked periodically by the customer and kept
functional.
Danger! Leaky connections can allow process medium, cleaning solution or control
(compressed) air to escape to the environment, presenting a hazard for persons and the
environment.
2.7.2 Heat protection
Caution! The sensor is not equipped with heat protection. During steam-sterilization
procedures in pipes, the surface of the sensor can reach high temperatures and cause
burns.
2.7.3 External influences
Notice: Objects falling on the sensor can damage or destroy the unit, or cause leaks etc.
Lateral forces may damage or destroy the unit.
2.8
Emergency measures
Notice: Always observe and comply with local regulations!