Problem
Only the "SYS" item is
visible in the "rF" menu.
Only the "SYS" and "Lrn"
items are visible in the
"rF" menu.
The items "APrt" and
"Time" are not visible in
the "rF" menu.
After calling up the
menu, the "rF" item is
not displayed.
The display "Er:X:11"
appears.
The display "Er:X:12"
appears.
The display "Er:X:16"
appears.
The Enter key
(send/print) is pressed
and the display "Er:X:71"
appears.
The Enter key
(send/print) is pressed
and the display
"Er:X:72" appears.
100 •
Cause / Solution
• The wireless module is deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 93).
• The wireless module is activated and no
wireless group is set up.
- Set up wireless group (see "Setting up a
wireless group (Lrn)" on page 93).
• No wireless printer registered to the wireless
group.
- Register the wireless printer to the wireless
group via the "Lrn" menu item (see "Setting
up a wireless group (Lrn)" on page 93).
• The scale's wireless module is defective.
- Inform seca service.
The scale has too high a load or too high a load in
one corner.
- Take the load off the scale or distribute the
weight more evenly.
- Restart the scale.
The scale has been switched on with too high a
load.
- Remove the load from the scale.
- Restart the scale.
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale.
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 93).
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a
wireless group (Lrn)" on page 93).