Guidelines and Manufacturer's Declaration – Electromagnetic Immunity
Frequency
Test
frequency
band
MHz
MHz
385
380 - 390
450
430 - 470
710
745
704 - 787
780
810
870
800 - 960
930
1 720
1 845
1 970
1 700 - 1 990
2 450
2 400 - 2 570
5 240
5 500
5 100 - 5 800
5 785
NOTE: If necessary, to achieve immunity test levels, the distance between the transmit antenna and the MiniScreen can
be reduced to 1 m. The 1 m test distance is permitted according to IEC 61000-4-3.
a
For some radio services, only the frequency for the radio link from the mobile communication device to the base
station (en: uplink) has been included in the table. The carrier must be modulated with a square wave signal with
50% duty cycle.
b
The carrier must be modulated with a square wave signal with 50% duty cycle.
c
As an alternative to Frequency Modulation (FM), pulse modulation at 50% duty cycle at 18 Hz may be used, as
this, if not the actual modulation, would be the worst case scenario
Table 4: Table 9 EN 60601-1-2, Test specifications for the immunity of enclosures to high-
Radio
a
Modulation
a
service
Pulse Modulation
TETRA 400
GMRS 460,
± 5 kHz Hub
FRS 460
LTE Band 13,
Pulse Modulation
17
GSM 800/900
TETRA 800,
Pulse Modulation
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
Pulse Modulation
DECT;
LTE Band 1, 3,
4, 25;
UMTS
Bluetooth,
WLAN 802.11
Pulse Modulation
b/g/n,
RFID 2450,
LTE Band 7
WLAN 802.11
Pulse Modulation
a/n
frequency wireless communication devices
Max. Perfor-
b
mance
W
b
1,8
18 Hz
c
FM
2
1 kHz Sinus
b
0,2
27 Hz
b
2
18 Hz
b
2
217 Hz
b
2
217 Hz
b
0,2
217 Hz
- 15 -
Distance
Immunity test level
m
V/m
0,3
27
0,3
28
0,3
9
0,3
28
0,3
28
0,3
28
0,3
9