Press and hold the
appears. Press
Sample type: Solid
Auxiliary Liquid: Water
Porous Material: On
Set the following parameters by pressing
Water Temperature
Weight
Oil Density
Measure the actual water temperature using a precision
thermometer. The balance calculates water density based on
the water temperature value entered.
The weight of the sample and the density of oil must be
measured in advance.
1. Weigh the oiled sample in the air with the balance and the
density determination kit.
2. When the * (symbol of stability) appears, press the
sample in the air.
1.888.610.7664
button until Density
to enter the Density Determination.
or
button to confirm the weight of the oiled
www.calcert.com