6.3. AFTER USE
•
Before disconnecting low pressure regulator from the medical pipe line
system terminal unit or medical regulator, reset fl ow control knob on
ZERO position.
•
Vent gas pressure from downstream equipment (i.e. via the connected
consumers).
•
Remove connections from user outlets.
•
Refi t protection caps. Before refi tting the caps ensure they are clean.
7. CLEANING
Remove general contamination with a soft cloth damped in oil free oxygen
compatible soap water & rinse with clean water.
Disinfections can be carried out with an alcohol-based solution (spray or
wipes).
If other cleaning solutions are used, check that they are not abrasive and
that they are compatible the product materials (including labels) and gas.
Do not use cleaning solutions containing ammonia!
Do not immerse in water or any liquid.
Do not expose to high temperature (such as autoclave).
8. MAINTENANCE
8.1. SERVICE AND PRODUCT LIFE TIME
SERVICE
GCE recommends that a product Periodic inspection is undertaken every
year to check proper functionality of the regulator. It should be done by
skilled technician.
It should not be assumed that Periodic Inspection recommended by GCE
covers every safety procedure or method required by local regulations or
statutory requirements, or that abnormal or unusual circumstances may
not warrant or suggest further requirements or additional procedures.
LIFE TIME AND WASTE MANAGEMENT
Maximum life time of the product is 10 years.
At the end of the product's life time, the product must be withdrawn from
service. The owner shall put in place a relevant procedure to ensure the
product cannot be used again.
The owner of the device shall prevent the reuse of the product and han-
dle the product in compliance with "Directive of European Parliament and
Council 2008/98/EC on waste".