Electrical connections
Power supply connection
➀ = Power supply cable
Fig. 1:
A = View A
Fig. 2:
Ground terminal for potential equalisation
Caution:
• The transmitter ➀ (compact and remote version) is to be securely connected to the
potential equalization system using the screw terminal ➁ on the outside of the trans-
mitter housing.
• With the remote version, the sensor's wiring compartment housing ➂ is to be
grounded using the external screw terminal ➃. Alternatively, the sensor can be con-
nected to the potential equalization system via the pipeline when a ground connection
according to regulations can be assured.
The table below contains the values that are identical for all versions, irrespective of the
type code.
Transmitter Promass 80
Terminals
1
L (+)
Power supply ➀
Designation
Functional
AC: U = 85...260 V
values
or
AC: U = 20...55 V
or
DC: U = 16...62 V
Power consumption:
15 VA / 15 W
Intrinsically
no
safe circuit
U
=
260 V AC
m
2
3
N (–)
Protective earth
Caution:
Follow ground
network requirements
for the facility.