Siemens SITRANS LCS050 Instructions De Service page 40

Table des Matières

Publicité

Désignation
Temperature unit
(TMP)
Device status acc. to
NE107
Device status
Counter for change
of parameters (PCO)
Actual electronics
temperature
Min. electronics
temperature
Max. electronics
temperature
Actual measuring cell
temperature
Min. measuring cell
temperature
Max. measuring cell
temperature
Actual resonance
frequency
Min. resonance
frequency
Max. resonance
frequency
Probe
Output
Output 2
Device name
Serial number
SITRANS LCS050
Instructions de service, 02/2021,
A5E50859888
ISDU (dez) ISDU
(hex)
291
0x0123
294
0x0126
295
0x0127
296
0x0128
297
0x0129
299
0x012B
300
0x012C
301
0x011C
302
0x011D
303
0x011E
304
0x0130
305
0x0131
306
0x0132
307
0x0133
308
0x0134
309
0x0135
310
0x0136
311
0x0137
Caractéristiques techniques et Dessins cotés
10.2 Communication d'appareil IO-Link
Taille
Type de
Entrée
(octet)
données
4
Float
RW
1
Unsigned8 RO
19
Unsigned16 RO
4
Unsigned32 RO
4
Float
RO
4
Float
RO
4
Float
RO
4
Float
RO
4
Float
RO
4
Float
RO
4
Float
RO
4
Float
RO
4
Float
RO
2
Unsigned16 RO
2
Unsigned16 RO
2
Unsigned16 RO
19
String
RO
16
String
RO
Plage de
valeurs
2 = FNO, 3=FNC
1001 = °C
1002 = °F
0 = Good
1 = Function
Check
2 =
Maintenance
required
3 = Out of
Specification
4 = Failure
-
-
-
-
-
-
-
-
-
-
-
0 = Not Covered
256 = Covered
512 = Covered
inside Window
768 = Covered
outside Window
0 = Open
1 = Closed
0 = Open
1 = Closed
-
-
33

Publicité