Error which has
occurred
t
Implausible
measured values.
4 TM IV_GA_GB.indd 109
4 TM IV_GA_GB.indd 109
Possible causes
If the battery symbol is
permanently illuminated
the batteries are dead
and have to be
replaced.
Device not positioned at
the level of the heart.
Cuff too big or too
small.
Cuff placed on top of
clothing.
Rolled-up clothing
impedes blood
circulation.
Cuff wrongly applied.
Cuff was not correctly
infl ated.
109
English
Remedy
Insert new batteries of
the same type (type AA
LR03).
Position the wrist at the
level of the heart, and
repeat the measurement.
Use an upper arm blood
pressure monitor.
Apply cuff on the naked
skin.
Wear loose clothing.
Make sure that rolled-up
sleeves do not impair
circulation in the arm.
Take note of the
instructions and images
showing how to apply
the cuff correctly to the
wrist.
Check the correct
position of the cuff.