Télécharger Imprimer la page

Beurer TL 85 Mode D'emploi page 139

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 24
Power frequen-
30 A/m
cy (50/60 Hz)
50Hz/60Hz
magnetic field
lEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Table 3:
Guidance and manufacturer's declaration – electromagnetic immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the
Immunity test
Conduced RF
150KHz to 80MHz
IEC61000-4-6
3Vrms
6Vrms (in ISM
and amateur radio
bands)
80% Am at 1kHz
Radiated RF
10V/m
IEC61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the [XXXXX] is used exceeds the
applicable RF compliance level above, the [XXXXX] should be observed to verify normal operation. If abnormal per-
formance is observed, additional measures may be necessary, such as re-orienting or relocating the [XXXXX].
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Device should assure that it is used in such an environment
IEC 60601
Compliance
test level
150KHz to 80MHz
T
3Vrms
6Vrms (in ISM
and amateur radio
bands)
80% Am at 1kHz
10V/m
30 A/m
50Hz/60Hz
level
Portable and mobile RF communications equip-
T
ment should be used no closer to any part of the
Device ,including cables, than the recommended
separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Recommended separation
distances:
d=1.2√P;
d=2√P
80MHz to 800MHz
d=1.2 P
800MHz to 2.7GHz
d=2.3 P
139
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hos-
pital environment.
Electromagnetic environment –
guidance
Where P is the maximum
T
output power rating of the
transmitter in watts (W)
T
according to the transmitter
manufacturer, d is the
recommended separation
distance in meters (m) Field
strengths from fixed RF
transmitters, as determined
by an electromagnetic site
surveya ,should be less
than the compliance level
in each
frequency range b
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
 

Publicité

loading