System Error Trouble Shooting Guide
Problem
Cause
No LEDs lit; no audible
Control unit may not
alarm heard.
be attached to a power
source.
The fuse may need
replacing.
Air-leak alarm
Pressure has fallen below
LED (6) lit permanently
the minimum operational
yellow with an audible
requirements.
alarm.
Blockage alarm
Airflow obstruction
LED (7) lit permanently
between the control unit
yellow with an audible
and the platform.
alarm.
Service LED (8)
Service required.
permanently lit amber.
Interface panel is lit
Interface panel is locked.
but unresponsive.
An audible alarm is
Loose or absent power
sounding; no LEDs lit.
lead.
Mains power accidentally
turned off.
Mains power failure.
Do not open the control unit. Opening the unit could cause personal injury or equipment damage.
NB. The Toto Touch control unit self calibrates on start up.
Initial Checks / Trouble Shooting
1.
Check that the mains power outlet is live and active.
2. Check the control unit is switched on.
3. If following points 1 and 2 does not resolve the
problem, switch off and unplug the control unit:
- Check the mains plug fuse (3A).
- Check the control unit fuse (1A).
4. Restart the control unit.
If the problem persists, contact Frontier Therapeutics
Ltd Tel. +44 (0)330 460 6030 for assistance.
1.
Press the alarm pause button.
2. Check the CPR valve is closed and correct.
3. Check all air cells and hoses for any air leakage.
4. Resolve and wait for the alarm to reset.
If the problem persists, contact Frontier Therapeutics
Ltd Tel. +44 (0)330 460 6030 for assistance.
1.
Press the alarm pause button.
2. Check for twists or kinks in the air hoses between
the platform and control unit.
3. Resolve and wait for the alarm to reset.
If the problem persists, contact Frontier Therapeutics
Ltd Tel. +44 (0)330 460 6030 for assistance.
Please contact Frontier Therapeutics Customer
Services for support Tel. +44 (0)330 460 6030
The Control Unit continues to function even though
the service LED remains lit.
1.
Check that screen lock LED is lit amber (9).
2. Press and hold the alarm mute button for four
seconds to deactivate.
If the problem persists, contact Frontier Therapeutics
Ltd Tel. +44 (0)330 460 6030 for assistance.
1.
Reconnect mains power lead to the control unit.
2. Check the mains power wall switch is turned on.
3. Attempt to restart the control unit.
If the problem persists, contact Frontier Therapeutics
Ltd Tel. +44 (0)330 460 6030 for assistance.
19
19
CPR Emergency Deflation System
The Toto system features a quick release CPR deflation valve which allows for rapid deflation for
emergency procedures.
In the event of an emergency, firmly pull and release the yellow tab.
To re-inflate the system, replace the CPR tag ensuring both sealing connectors are firmly attached
and restart the Toto Touch control unit.
CPR Deflation Valve Closed
CPR Deflation Valve Open
6
8
7
9
20
20