3
53220A/53230A Measurements
– channel_pair settings are (@1),(@2)|(@2),(@1)|(@1),(@3)|
Frequency ratio examples
// using MEASure? - ratio measurement expecting a 1:1
// ratio, set 6 digits of resolution of ratio measurement
*RST
MEAS:FREQ:RAT? 1, 1.0E-6, (@1),(@2)
-------------------------------------------------------
// using CONFigure - ratio measurement expecting a 1:1
// ratio, set 9 digits of ratio resolution
*RST
CONF:FREQ:RAT 1, 1.0E-9, (@1),(@2)
READ?
Notes
1 The digits of resolution in the above ratio measurements is set by the
2 See
96
(@3),(@1)|(@2),(@3)|(@3),(@2). Within the pair, the first channel represents
the ratio's numerator and the second channel represents the ratio's
denominator. The default channel_pair is (@1),(@2).
// reset to start from known state
// reset to start from known state
INP:LEV 1.5
// set a 1.5V threshold level (ch. 1)
INP2:LEV 1.5
// set a 1.5V threshold level (ch. 2)
resolution parameter (1.0E-6, 1.0E-9). In effect, the exponent corresponds
to the number of digits. See
information on the relationship between gate time and reading resolution.
Chapter 4
for additional information on counter threshold levels and on
configuring the input signal path.
Frequency measurements