A
53220A/53230A Error Messages
Table A-1
53220A/53230A error message descriptions. (continued)
Code
Message
-178
Expression data not allowed
Execution errors
-203
Command Protected
-213
INIT ignored
-221
Settings conflict
-222
Data out of range
-223
Too much data
-224
Illegal Parameter Value
Out of memory;
-225
measurement data overrun
-230
Data corrupt or stale
-240
Hardware error
-241
Hardware missing
Mass storage error:
-250
file read/write error
-252
Missing Media
-254
Media full
File or folder name not found The folder or file name specified in command does not exist.
-256
288
Description
An expression was encountered at an instance where it is not allowed by
the instrument.
A password-protected command or query could not be executed
because the command was disabled.
INITiate:IMMediate was received while the counter is currently
initiated.
The statement appended to the 'Settings conflict' message describes
the conflict and how it was resolved.
The parameter value received was outside the range for the given
command.
The block, element, or string parameter received is valid, but memory
limitations prevent command execution.
An exact value from a list of choices was expected.
Internal buffers are full as data is taken at a faster rate than can be
processed. Decreasing the reading rate (gate time) can prevent this
error.
Attempting to retrieve data following a reset or change to the counter
configuration.
Most hardware errors are detected at instrument power-on. If a
hardware error occurs, cycle power again. If the error remains, contact
Keysight.
The function specified requires optional channel 3. The channel is either
missing or incorrectly installed.
An error occurred reading from, or writing to a file in internal flash
memory or on a USB memory device. File space may have been
exceeded or the USB device removed before the operation was
complete.
Internal flash memory or the USB device was not found during a file
operation.
Internal flash memory or the USB device has no memory space
available to create the folder or file, or to perform the file operation.
Keysight 53220A/53230A User's Guide