Triggering and Gating
Gate source ADVanced
Gate source ADVanced enables extended control of the gate signal through the
counter's SENSe:GATE commands (see
Advanced gate control - gate stop hold off and gate
Continuous totalizing
Continuous totalizing as configured by the command:
CONFigure:TOTalize:CONTinuous
sets the gate source to TIME and the gate time to INFinity. The command also
sets the input threshold to 0.0 V and the edge (events that are totalized) to
positive. Refer to
information on the commands in the INPut subsystem used to change these
parameters.
Reading the current count
gate times, the current count can be read with the command:
[SENSe:]TOTalize:DATA?
For example:
CONF:TOT:CONT
INIT
SENS:TOT:DATA?
180
Chapter 4, "53220A/53230A Input Signal Conditioning"
During continuous or timed totalizing with long
// configure continuous totalizing
// start measurement
// query current count
Advanced gate control - gate start
stop).