3
53220A/53230A Measurements
– channel is optional channel 3 specified as (@3).
Carrier frequency measurement examples
// measure the carrier frequency using MEASure? and its
// default settings
*RST
MEAS:FREQ:BURS? (@3)
-----------------------------------------------------
// measure the carrier frequency using CONFigure, set
// automatic gate configuration
*RST
CONF:FREQ:BURS (@3)
READ?
Notes
1 See
2 See
126
// reset to start from known state
INP3:BURS:LEV -6
SENS:FREQ:BURS:GATE:NARR OFF // narrow pulse mode off
SENS:FREQ:BURS:GATE:AUTO ON
Burst measurement detector threshold
setting the threshold for pulse measurements.
Burst carrier frequency measurements
narrow pulse mode and on automatic and manual gate control for carrier
measurements.
// reset to start from known state
// set a detector threshold of -6 dB
// automatic gate set up
Keysight 53220A/53230A User's Guide
in
Chapter 4
for information on
in
Chapter 5
for information on