Télécharger Imprimer la page

Keysight 53220A Mode D'emploi page 217

Masquer les pouces Voir aussi pour 53220A:

Publicité

Les langues disponibles

Les langues disponibles

CALCulate[1]:AVERage:PTPeak?
– returns the average peak-to-peak value (average maximum value - average
Example: Computing individual statistics
The following example shows these reading parameters queried individually.
CONF:PER (@1)
INIT
*WAI
CALC:AVER:MIN?
CALC:AVER:MAX?
CALC:AVER:AVER?
CALC:AVER:PTP?
Allan deviation
Allan deviation which is used to estimate stability, is selectable from the front
panel
CALCulate[1]:AVERage:ADEViation?
Allan deviation is used for frequency and period measurements only, and requires
continuous (gap-free) measurements (SENSe:FREQuency:MODE CONTinuous) for
highest accuracy. Continuous mode is available with the 53230A only.
Allan deviation is also available with the 53220A; however, the 53220A does not
support continuous (gap free) measurements.
Keysight 53220A/53230A User's Guide
minimum value) of all readings in the reading count.
TRIG:COUN 2
SAMP:COUN 100
SENS:FREQ:GATE:TIME 10e-3
CALC:STAT ON
CALC:AVER:STAT ON
(Figure
6-3) or can be enabled with the command:
53220A/53230A Math, Graphing, and Data Logging
// configure period measurement
// trigger count = 2
// 100 rdgs per trigger
// 10 ms gate time
// enable CALCulate1 subsystem
// enable statistics
// initiate the measurements
// wait for all rdgs to complete
// query individual parameters
6