shown first sequentially
set last programmatically
SENSe:FREQuency:GATE:SOURce
SENSe:TOTalize:GATE:SOURce
SENSe:TINTerval:GATE:SOURce
* not available with time interval
SENSe:GATE:STARt:SOURce
SENSe:GATE:EXTernal:SOURce
SENSe:GATE:STARt:SLOPe
SENSe:GATE:STARt:DELay:SOURce
SENSe:GATE:STARt:DELay:EVENts
SENSe:GATE:STARt:DELay:TIME
SENSe:GATE:STOP:HOLDoff:SOURce
SENSe:GATE:STOP:HOLDoff:EVENts
SENSe:GATE:STOP:HOLDoff:TIME
SENSe:GATE:STOP:SOURce
SENSe:GATE:EXTernal:SOURce
SENSe:GATE:STOP:SLOPe
Gate stop source does not apply to rise/fall
time, duty cycle, pulse width, phase, and
single-period measurements
Figure 5-2
Keysight 53220A/53230A User's Guide
(from Triggered State)
Gate Source
ADVanced
Gate Start Source
EXTernal
External Gate Source
BNC
INPut[1] | INPut2 *
Gate Slope
Gate Delay
(input channel)
Gate Start (open)
Gate Stop Hold Off
(input channel)
Gate Stop Source
EXTernal
External Gate Source
EXTernal /
INPut[1]* | INPut2 *
Gate Slope
Gate Stop (closed)
Trigger and gate cycle - gating (cont'd)
TIME
IMMediate
EXTernal
INPut[1]* | INPut2 *
IMMediate
(internal)
SENSe:FREQuency:GATE:TIME
SENSe:TOTalize:GATE:TIME
SENSe:TINTerval:GATE:SOURce IMMediate
IMMediate
(delay disabled)
(for all EXTernal Sources)
Gate In/Out BNC Disable
OUTPut:STATe OFF
IMMediate
(holdoff disabled)
IMMediate
(internal)
(measurement completes on next edge of
input signal after gate is closed)
auto gate extension
(frequency and period measurements)
Triggering and Gating
(to SAMPle COUNt reached?)
Gate Start (open)
Gate Stop (closed)
Gate Start (open)
SENSe:FREQuency:GATE:POLarity
SENSe:TOTalize:GATE:POLarity
SENSe:TINTerval:GATE:POLarity
INPut[{1|2}]:LEVel[{1|2}][:ABSolute]
Gate Stop (closed)
5
159