QUICK UNLOADING OF THE TIPS
To quickly empty tips filled with reagent, raise the loading lever.
PARTIAL TIP FILLING
When necessary, the tips can be partially filled by depressing the loading lever part of the
way down.
The Multistepper module should be stored with the filling lever in the upmost position. This
is essential to ensure proper function of the pipette.
However, if MULTISTEPPER MODULE is stored for long periods, the pistons may jam.
When that happens and pistons are about 1 mm out of the tip cones, press the tip cones
firmly against an even surface, in other case press the filling lever until piston are released.
(Fig. 4).
The Multistepper module requires a limited amount of maintenance and service. If used
daily, clean the cylinders and O-ring base with a napless cloth once or twice a year.
Similarly the O-rings and pistons should be lightly greased with the lubricant grease
provided with the pipette.
There is no need to adjust the module.
In the case of malfunction, please return the Multistepper module to the nearest
Labsystems subsidiary or authorized agent.
If harmful substances have been handled with the Multistepper, please clean the module
thoroughly before shipping it for service.
Weight
Spacing
Channels
Volume Uptake
Volume Delivery
STORAGE
SERVICE
SPECIFICATIONS
162 g
9 mm
8
1250 µl/channel (maximum)
24x50 µl/channel
12x100 µl/channel
8 x150 µl/channel
6x200 µl/channel
5x250 µl/channel
7