104
1.3
Technical Specifications
Titrator TitroLine
1.3.1
Translation of the legally binding German version
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1: 2013
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: 2010 for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
All common titration solutions.
As reagent water and all non-aggressive non-organic and organic fluids are allowed.
If using combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
For fluids with higher viscosity ( 5 mm
speed can be adjusted.
Fluids with viscosity over 20mm
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
®
7500 KF to "warm up".
TitroLine
Measuring input (μA): Karl Fischer (Dead-Stop) connector (µA) for double platinum electrode.
Measurement range I [A]
Display:
3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Measurement input μA: (Dead-Stop-) connector for double platinum electrode
Input:
Power supply:
power supply 90-240 V; 50/60 Hz, power input: 30 VA
RS-232-C Interface:
separated galvanically through photocoupler, Daisy Chain function available
Data bits:
Stop bit:
Start bit:
Parity:
Baud rate:
Address:
RS-232-1
for computer, input Daisy Chain
RS-232-2
devices of SI Analytics
- Titrator TitroLine
- Sample Changer TW alpha plus, TW 7400
- Piston burette TITRONIC
- Balances of the types Mettler, Sartorius, Kern, Ohaus, (for more, please contact us)
- Exit Daisy-Chain
* The measurement uncertainty of the sensor probe has to be taken into account as well.
®
7500 KF
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
Polarisation voltage variably adjustable from 40 ... 220 mV.
Connector: 2 x 4 mm – sockets.
100
50
10
5
(Connection sockets: 2 x 4 mm)
Use the power supply TZ 1853 only!
adjustable, 7 or 8 Bit (default: 8 Bit)
adjustable, 1 or 2 Bit (default: 1 Bit)
static 1 Bit
adjustable: even / odd / none
adjustable: 1200, 2400, 4800, 9600, 19200 (Default 4800 baud)
adjustable, (0 to 15, default: 01)
®
:
®
7000 / 7500 / 7500 KF / 7750 / 7800
®
Display resolution
Measurement accuracy* without sensor probe
0.1
0.1
0.1
0.1
300 and 500, TITRONIC
(Release: 21. February 2020)
-5 /+ 3 A 1 Digit
+/- 3 A 1 Digit
+/- 1 A 1 Digit
+/- 0.2 A 1 Digit
®
®
110 plus, TITRONIC
universal,