Canister Removal And Replacement; Leaks - KCI V.A.C.VIA Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CANISTER REMOVAL AND REPLACEMENT

1. Turn therapy off.
2. Slide dressing tubing clamp close to where tubing plugs into canister. Close clamp.
3. Unplug tubing from canister tubing ports.
4. Remove therapy unit from carrying case, if in use.
5. Press tab on canister to remove used canister from therapy unit (Fig. 14).
6. Install new canister (see V.A.C.VIA™ Canister Installation section).
7. Return therapy unit to carrying case if desired.
8. Reattach dressing tubing to canister tubing ports.
9. Release tubing clamp.
10. Turn therapy on.
NOTE: Dispose of used canister according to institution and local environmental regulations.

LEAKS

When the therapy unit detects a significant leak, a visual and audible leak alarm will activate (See Alarms
section).
During a leak condition, a solid yellow LED above the leak symbol will turn on. The system will sound
two beeps repeating every 15 seconds. The yellow LED will remain solid until the leak condition has
been corrected, at which time it will turn off. The therapy unit will continue the alarm until condition is
corrected. When leak condition has been corrected, audible alarms will stop, and visual alarms will turn
off.
NOTE: The green LED next to the therapy pressure selector button, 75 mmHg or 125 mmHg, will remain on
during the leak condition.
24

Publicité

Table des Matières
loading

Table des Matières