Technical Information; Error Messages - uebe visomat comfort form Mode D'emploi

Masquer les pouces Voir aussi pour visomat comfort form:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

E

1. Error messages

Error
Possible cause
Display Err - 300 Excessive cuff pressure.
The cuff was inflated to
its maximum due to the
movement of the arm
or body.
The air hose may be
kinked or the air plug
blocked.
Display Err - 1
Measurement error;
measurement could not
be carried out due to
movement or speaking.
The acquired values lie
outside the measuring
range of the device (see
„Technical data" on
page 60).
Display Err - 2
Inflation error; pressure
built up too quickly or
too slowly.
Fault has been iden-
tified. The cuff is not
correctly attached
Display Err - 3
Deflation rate too high
or too low.
Display 0
Start/Stop button was
0
accidentally pressed
while the batteries were
being replaced or while
the mains adapter was
being plugged in.
Arrhythmia, movement
disorders, trembling,
wobbling (objects), deep
breathing etc.
EN-58

Technical information

Remedy
- Repeat the measurement
- Do not move your arm
- Do not speak
Inspect the hose line and air
plug for blockages
- Repeat the measurement
- Do not move your arm
- Do not speak
Repeat the measurement
Check that the air connector
is fitted properly, repeat
measurement.
- Repeat the measurement
- Do not move your arm
- Do not speak
Call the Service hotline if
this occurs repeatedly.
Switch unit off and then on
again using the Start/Stop
button.
Repeat measurement after
resting for 3-5 minutes.
Listen out for the device's
pulse beep signals. Possible
influences on the measure-
ment results (depending on
the severity of the arrhyth-
mia) must be discussed with
your doctor!
Technical information
Error
Possible cause
The motion control
detected movement du-
ring the measurement.
Measured
Too little rest before ta-
values/pulse
king the measurement.
values are
Movement or talking
extremely high
during the measure-
or low.
ment.
Incorrect measurement
posture
Measurements
Blood pressure is not a
taken in direct
fixed value. It can go up
succession
or down by more than
produce diffe-
20 mmHg within a few
rent measured
minutes.
values.
Measurement
Device detects a fault or
was interrupted,
that the diastolic value
the cuff deflated
is too low.
to 0 and the
Movement during mea-
measurement
surement
restarted.
No display
Batteries inserted
after device is
incorrectly
switched on
Batteries flat
Battery contacts dirty.
Dashes in
No measured values
display after
have been stored.
pressing Memo-
ry button
Measurement is
Batteries flat
interrupted
The time is
Batteries flat
visible in the
display, but the
measurement is
interrupted and
the symbol
appears.
E
Remedy
- Repeat the measurement
- Do not move your arm
- Do not speak
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 45.
Repeat the measurement
and maintain the correct
posture. (Page 49)
In order to obtain compara-
ble results, always carry out
the measurements under
the same conditions.
Device shows no fault.
Repeat measurement
If necessary, interrupt mea-
surement and repeat after 5
minutes' rest.
Check position of batteries.
Replace batteries.
Clean battery contacts with
a dry cloth.
Repeat measurement
Replace batteries.
Replace batteries.
EN-59