1. Failure and error messages

Failure
Possible cause
encountered
Display Err - 300
Excessive cuff pressure. Cuff
inflated to maximum due to
movement of the arm or body.
The hose may be kinked or the air
connector blocked.
Display Err - 1
Measurement error. Measurement
could not be carried out due to
movement of the arm or body.
Display Err - 2
Inflation error, pressure built up too
quickly or too slowly.
Fault was recognised. Cuff is not
correctly fitted.
Display 0
Start/Stop button accidentally
0
pressed while battery was being
changed.
Irregular heartbeat, disruptive
movements, shaking, wobbling
(objects), breathing in deeply, etc.
EN-46

Technical information

Corrective action
- Repeat measurement
- Do not move arm
- Do not talk
- Check that hose with air
connector is clear (unobstructed)
- Repeat measurement
- Do not move arm
- Do not talk
Check seating of air connector,
repeat measurement.
- Repeat measurement
- Do not move arm
- Do not talk
Switch unit off and then on
again with Start/Stop button.
Repeat measurement after 3-5
minutes rest. Possible factors
(depending on the severity of
the arrhythmia) affecting the
measurement results must be
discussed with your doctor.
Technical information
Failure
Possible cause
encountered
Measured values
Too little rest before the
/ pulse values are
measurement. Movement or
extremely high
talking during measurement.
or low
Has the proper cuff been selected?
Cuffs that are too small result in
excessive blood pressure values.
Incorrect measurement position.
Consecutive
Blood pressure is not a fixed
measurements
quantity. It may vary in the upward
produce different
or downward direction within
measured values
a few minutes by more than 20
(readings).
mmHg.
Measurement was
Strong pulse restlessness or
interrupted and cuff
movement detected, unit rejects
first deflated and
measurement
then inflated
No display after unit
Batteries inserted incorrectly
is switched on
Batteries flat
Battery contacts dirty.
Dashes in display
No measured values stored.
after Memory button
is pressed
Batteries flat
Measurement
Batteries flat
interrupted
Date/time not shown
- Function accidentally deactivated
in display
- Function was not reactivated
after battery replacement
E
Corrective action
Repeat measurement after
a break of approx. 3-5 mins.
Observe instructions for use
P. EN-36.
Measure your arm
circumference in the middle of
the upper arm and compare
this with the details given on
the cuff.
Check conditions and repeat
measurement. Observe
instructions for use P. EN-36.
---------
If necessary, interrupt
measurement, measure again
after a break of 5 minutes.
Check position of batteries.
Change batteries.
Clean battery contacts.
No display when no measured
values are stored.
Change batteries.
Change batteries.
Activate date and time as
described on P. EN-38.