1. Failure and error messages

Failure
Possible cause
encountered
Display Err
Excessive cuff pressure. Cuff inflated
- 300
to maximum due to movement of
the arm or body.
The hose may be kinked or the air
connector blocked.
Display Err - 1
Measurement error. Measurement
could not be carried out due to
movement of the arm or body.
Display Err - 2
Inflation error, pressure built up too
quickly or too slowly.
Fault was recognised.
Irregular heartbeat, disruptive
movements, shaking, wobbling
(objects), breathing in deeply, etc.
Unusual
Movement or talking during
measured
measurement, resting time not
values
observed, feet possibly crossed,
smoking or coffee cunsumption.
EN-46
Technical information
Corrective action
- Repeat measurement
- Do not move arm
- Do not talk
- Check that hose with air
connector is clear (unobstructed)
- Repeat measurement
- Do not move arm
- Do not talk
Check seating of air connector,
repeat measurement.
- Repeat measurement
- Do not move arm
- Do not talk
Repeat measurement after 3-5
minutes rest. Possible factors
(depending on the severity of
the arrhythmia) affecting the
measurement results must be
discussed with your doctor.
Check conditions and repeat
measurement. Observe
instructions for use P. EN-36.
Technical information
Failure
Possible cause
encountered
Display 0
Start/Stop button accidentally
0
pressed while battery was
being changed.
Measurement was
Strong pulse restlessness or
interrupted and
movement detected, unit
cuff first deflated
rejects measurement
and then inflated
Movement during
measurement
No display after
Batteries inserted incorrectly?
unit is switched on
Batteries flat?
Battery compartment dirty?
Dashes in display
No measured values stored.
after Memory
button is pressed
Batteries flat?
Measurement
Batteries flat?
interrupted
Measured values
Was the correct measurement
extremely high
position maintained?
or low
Was the necessary
rest observed prior to
measurement?
D
Corrective action
Switch unit off and then on again
with Start/Stop button.
If necessary, interrupt
measurement, measure again after
a break of 5 minutes.
If necessary, interrupt
measurement, measure again after
a break of 5 minutes.
Check position of batteries.
Change batteries.
Clean battery compartment.
No display when no measured
values are stored.
Change batteries.
Change batteries.
Repeat measurement, ensuring
correct body posture. Do not move
your arm and do not speak.
Repeat measurement after a break
of approx. 3-5 mins.
- Do not move arm
- Do not talk
EN-47