1. Failure and error messages

Failure
Possible cause
encountered
Display Err -
Excessive cuff pressure. Cuff
300
inflated to maximum due to
movement of the arm or body.
The hose may be kinked or the
air connector blocked.
Display Err - 1
Measurement error.
Measurement could not be
carried out due to movement
of the arm or body.
Display Err - 2
Inflation error, pressure built
up too quickly or too slowly.
Fault was recognised.
Display Err - 3
Deflation rate too high or
too low.
Irregular heartbeat, disruptive
movements, shaking,
wobbling (objects), breathing
in deeply, etc.
Unusual
Movement or talking during
measured
measurement, resting time
values
not observed, feet possibly
crossed, smoking or coffee
cunsumption.
EN-46
Technical information
Corrective action
- Repeat measurement
- Do not move arm
- Do not talk
- Check that hose with
air connector is clear
(unobstructed)
- Repeat measurement
- Do not move arm
- Do not talk
Check sitting of air connector,
repeat measurement.
- Repeat measurement
- Do not move arm
- Do not talk
Call the Service hotline if this
occurs repeatedly.
Repeat measurement after 3-5
minutes rest. Possible factors
(depending on the severity of
the arrhythmia) affecting the
measurement results must be
discussed with your doctor!
Check conditions and repeat
measurement. Observe
instructions for use P. EN-36.
Technical information
Failure
Possible cause
encountered
Display 0
Start/Stop button
0
accidentally pressed while
battery was being changed.
Measurement
Strong pulse restlessness or
was interrupted
movement detected, unit
and cuff first
rejects measurement
deflated and
Movement during
then inflated
measurement
No display after
Batteries inserted
unit is switched
incorrectly?
on
Batteries flat?
Battery compartment dirty?
Measurement
Batteries flat?
interrupted
Measured
Was the correct
values
measurement position
extremely high
maintained?
or low
Was the necessary
rest observed prior to
measurement?
D
Corrective action
Switch unit off and then on
again with Start/Stop button.
If necessary, interrupt
measurement, measure again
after a break of 5 minutes.
If necessary, interrupt
measurement, measure again
after a break of 5 minutes.
Check position of batteries.
Change batteries.
Clean battery compartment.
Change batteries.
Repeat measurement,
ensuring correct body posture.
Do not move your arm and do
not speak.
Repeat measurement after a
break of approx. 3-5 mins.
- Do not move arm
- Do not talk
EN-47