Time Adjustment - Topcom BPM ARM 5500 XLD Manuel D'utilisateur

Masquer les pouces Voir aussi pour BPM ARM 5500 XLD:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 25
10.1

Time adjustment

PUL/min
11
Troubleshoot
Error message
Reason
on display
Batteries low
7
Er0
Pressure system is unstable before
measurement.
Er1
Systolic pressure detect fail
Er2
Diastolic pressure detect fail
Er3
Pneumatic system blocked or cuff is
too tight during inflation
Er4
Pneumatic system leakage or cuff is
too loose during inflation
Er5
Inflation above 300mmHg
Er6
Over 3 minutes while pressure
above 15 mmHg
Er7
EEPROM accessing error
Er8
Device parameter check error
Er9
MCU self-verity error
ErA
Span error: pressure sensor
parameter error
All manuals and user guides at all-guides.com
To adjust the date an time:
Press the start/stop
simultaneously. The month will blink on the display.
Press the memory button
the month.
Press the start/stop button
The day will blink on the display.
Change the day, hours and minutes as described above,
using the memory button
stop button
After setting the minutes, the device is ready to use.
TOPCOM BPM Arm 5500 XLD
and memory button
9
M
successively to change
8
to confirm the month.
9
to change and the start/
M
8
to confirm every setting.
9
Solution
Change the batteries
Don't move and try again.
Place the cuff correctly and
try again.
Measure again after 5
minutes. If the monitor is still
abnormal, please contact the
local distributor or the
factory.
11

Publicité

Table des Matières
loading

Table des Matières