Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emissions
The MODEL IH 55 is compliance for each EMISSIONS test specified by the standard, e.g. EMISSIONS class and group.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The MODEL IH 55 is compliance for each IMMUNITY test specified by the standard, e.g. IMMUNITY test level.
Immunity test
Electrostatic discharge (ESD) IEC
61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Electrical fast transient/burst IEC
61000-4-4
Surge
IEC 61000-4-5
Conducted disturbances induced by
RF fields
IEC 61000-4-6
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
NOTE a) U
is the a.c. mains voltage prior to application of the test level;
T
All manuals and user guides at all-guides.com
Compliance
Electromagnetic environment – guidance
Group 1
The MODEL IH 55 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The MODEL IH 55 is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies
Class A
buildings used for domestic purposes.
Complies
IEC 60601-1-2 test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 KV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
±2 kV
100 kHz repetition frequency
±0.5 kV, ±1 kV line-to-line;
±0.5 kV, ±1 kV and ±2 kV line-to-ground;
3V
0.15MHz-80MHz
6V in ISM and amateur radio bands between 0.15MHz
and 80MHz
80% AM at 1kHz
0% U
: 0.5 cycle a)
T
At 0o, 45o, 90o, 135o, 180o, 225o, 270o, and 315o.
0% U
: 1 cycle
T
70% U
: 25/30 cycles
T
Single phase: at 0°
0% U
: 250/300 cycles
T
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 KV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
±2 kV
100 kHz repetition frequency
±0.5 kV, ±1 kV line-to-line;
±0.5 kV, ±1 kV, and ±2 kV line-to-ground;
3V
0.15MHz-80MHz
6V in ISM and amateur radio bands between 0.15MHz
and 80MHz
80% AM at 1kHz
0% U
: 0.5 cycle a)
T
At 0o, 45o, 90o, 135o, 180o, 225o, 270o, and 315o.
0% U
: 1 cycle
T
70% U
: 25/30 cycles
T
Single phase: at 0°
0% U
: 250/300 cycles
T
87