Commands:
NAMUR Commands
Function
IN_NAME
Input description name
IN_PV_X
Reading the real value
X=1;2;3;4;
IN_SOFTWARE
Input software ID number date, version
IN_SP_X
Reading the set rated value
X=1;2;3;4;6;
IN_TYPE
Input laboratory unit ID
OUT_NAME
Output description name. (Max. 10 characters, default: LR 1000)
OUT_SP_12@n
Setting the WD safety temperature with the echo of the set value
OUT_SP_42@n
Setting the WD safety speed with the echo of the set value
OUT_SP_X n
Setting the rated value to n
X=1;2;4;6
OUT_WD1@m
Watchdog mode 1: When a WD1 event occurs, the heating and Stirring functions are shut
down and message PC 1 is displayed. Set the watchdog time to m (20...1500) seconds, with
echo of the watchdog time. This instruction starts the watchdog function and must be sent
within the set watchdog time.
OUT_WD2@m
Watchdog mode 2: When a WD2 event occurs, the speed setpoint will be set to the WD safety
setpoint speed and the temperature setpoint will be set to the WD safety setpoint temperature.
The PC 2 warning is displayed. The WD2 event can be reset with OUT_WD2@0-resetting also
blocks the watchdog function. Set the watchdog time to m (20...1500) seconds, with echo of
the watchdog time. This command starts the watchdog function and must be sent within the
set watchdog time.
RESET
Switching off the instrument function.
START_X
Starting the instrument's (remote) function
X=1;2;4
STATUS
Display of status
1S: mode of operation A
2S: mode of operation B
3S: mode of operation C
S0: manual operation without fault
S1: Automatic operation Start (without fault)
S2: Automatic operation Start (without fault)
STOP_X
Switching off the instrument - (remote) function
X=1;2;4
Variables set with OUT_SP_X are maintained.
46
All manuals and user guides at all-guides.com
<0: error code: (-1)
- 1: error 1
- ... (see table)
-31: error 31
-83:wrong parity
-84: unknown instruction
-85:wrong instruction sequence
-86: invalid rated value
-87: not sufficient storage space