Chapter II
Number of counts :
Range selection :
Additional error according to crest factor : 0.2 % for a crest of 2 to 3
(Specification given full scale for a squarewave signal pulse 200 µs wide)
Additional error in IAC+DC with a direct current at input : 1 %
5.5.
Resistance / Continuity
Ranges
500 /
5 k
50 k
500 k
5 M
50 M
Overload protection can be reset automatically.
It is highly recommended to use very short and shielded test leads for measurements in
this range (> 1 M).
Number of counts :
Range selection :
Maximum open circuit voltage :
Detection threshold in continuity mode : 10 to 20
Response time in continuity mode :
5.6.
Capacitance
Discharge all capacitors before performing measurements.
Ranges
50 nF
500 nF
5 µF
50 µF
500 µF
5000 µF
50 mF
Overload protection can be reset automatically.
It is highly recommended to use very short and shielded test leads for measurements in
this range.
Number of counts :
Range selection :
Maximum open circuit voltage :
Caution !
Do not perform capacitance measurements in dangerous zones.
Intrinsically safe portable digital multimeter
Accuracy
0.07% R + 8 D
0.07% R + 5 D
0.07% R + 2 D
0.3% R + 2 D
1% R + 2 D
Measurement
Accuracy
current
100 nA
1 µA
10 µA
1% R + 2 D
100 µA
1 mA
50 000 (or 5 000 - refer to §. 3.4.)
automatic or manual for the 500 µA, 5 mA, 50 mA,
500 mA ranges
0.5 % for a crest of 3 to 6
Measurement
Protection
current
1 mA
100 µA
10 µA
1 µA
100 nA
50 nA
50 000 (or 5 000 - refer to §. 3.4.)
automatic or manual (fixed in continuity mode)
7 V
1 ms
Max
measurement
time
0.5 s
1.5 s
3 s/mF
5 000
automatic or manual
7 V
X02137E00 - Ed. 1 - 02/17
Resolution
10 m
100 m
1
600 V
RMS
10
100
1 k
Protection
Resolution
10 pF
100 pF
600 V
10 nF
RMS
100 nF
10 µF
1 nF
1 µF
39