• Water way connection
Water pipe
①
Water
entrance
③
• Sketch map for connection of water supply system
Nipple
Seal
Handpiece
The connector of cable
Cable
Cap of
connector
②
④
4
• Sketch map for connection of detachable handpiece
Torque wrench
Endochuck
fig 3
1. Scaling function (G) and periodontal function (P)
• Open the packing box, make sure that all the parts and accessories are complete according to the packing list. Take the main
unit out of the box and put it on a stable plane.
• Turn the water control know to the max.
• Insert the plug of the foot switch to its socket (See picture 2).
• Connect one end of the water pipe to the water entrance of the main unit, and the other end to the water source (See picture 3).
• Connect the handpiece and the connector of cable correctly.
• Insert the plug of the power source to its socket.
• Switch on the main unit.
- For the scaling function (G) the scaling indicator and the first five power indicators shine.
- For periodontal function (P) the scaling indicator and the first three power indicators shine.
• Select a suitable scaling tip as you need then screw it on the handpiece tightly by the torque wrench (See picture 5).
• Under the normal working state of the scaling tip, light touch and a certain to-and-fro motion will eliminate the tartar without
heating. Overexertion and long-time lingering are forbidden.
• Vibrating intensity: Adjust the power of the vibrations as you need. According to patient's different sensitivity and the rigidity of
the tartar, adjust the vibrating intensity during the clinical treatment.
• Water volume adjust: Step on the foot switch, and the tip begins to vibrate, then turn the water control knob to form fine spray
to cool down the handpiece and clean the teeth.
• The handpiece must be handled in the same gesture as a pen in hand.
• During the clinical treatment, be sure not to make the end of tip touch the teeth vertically and not to make the tip overexert in
case of leading to a pain and damaging the tip.
• After finishing operation, keep the machine working for 30 seconds on the water supply condition in order to clean the hand-
piece and the scaling tip.
• Unscrew the scaling tip and pull out handpiece, then sterilize them.
Notice: Don't pull out the handpiece when the foot switch is stepped on and the machine is producing ultrasonic vibration.
fig 4
Tip
Handpiece
Endo wrench
File
Screw on
2. INSTALLATION AND INSTRUCTION FOR USE
5
GB
Screw on
Screw on
fig 5