Télécharger Imprimer la page

Precautions; Notice When Using Equipment; Contra-Indications; Storage And Maintenance - CFPM TRI-SCALER Compact Manuel D'utilisation

Masquer les pouces Voir aussi pour TRI-SCALER Compact:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
Remarks
Remark 1.
Turn the water control knob as symbol showed.
Remark 2.
Clean the water pipe with the multi-function syringe of the dental unit (See Figure 10):
Multi-function syringe
• Cut the water pipe at a distance of 10cm to 20cm from the water entrance.
• Get through to the power and turn on the power switch.
• Approach the multi-function syringe of dental unit to the water pipe.
• Disassemble the tip or handpiece.
• Step on the foot switch.
• Turn on the switch of the multi-function syringe, press the water into the machine and the impurity blocked in the water pipe
can be eliminated.
Remark 3.
If the scaling tip has been screwed on tightly and there is fine spray too, the following phenomenon show that the scaling tip is
damaged:
- The vibrating intensity and the water spray degree become weak obviously.
- During treatment, it produces the sound like "buzz" from the scaling tip.

1. Notice when using equipment

1. Keep the scaler clean before and after operation.
2. The handpiece, scaling tip, torque wrench, endo wrench and endochuck must be sterilized before each treatment.
3. Don't screw or unscrew the scaling tip and endochuck when stepping on the foot switch.
4. The scaling tip must be fastened and there must be fine spray or drip coming out from the tip when you use it.
5. Change a new one when the tip and ultrasonic file are damaged or worn excessively.
6. Don't twist or rub the tip and endochuck.
7. Don't use impure water source and be sure not to use normal brine instead of pure water source.
8. If use the water source without hydrolic pressure, the water surface should be one meter higher than the head of the patient.
9. Ensure the connector of handpiece and the socket of the cable dry before installing the handpiece.
10. Don't pull the cable forcibly in case of the handpiece falling off from the cable.
11. Don't knock or rub the handpiece.
12. After operation, turn off power, and pull out the plug.
13. We are only responsible for the satety on the following conditions:
- The maintenance, repair and modification are made by the manufacturer or the authorized dealer.
- The changed components are original of "DTE" and are operated according to instruction manual.
14. The internal screw thread of the scaling tips produced by some manufacturers maybe coarse, rusty and collapsed. This will
damage the external screw threat of the handpiece irretrievably.

2. Contra-indications

This equipment mustn't be used on hemophilia disease patients.
The patients or doctors with heart pacemaker are forbidden to use this equipment.
The heart disease patient, pregnant woman and children should be cautious to use the equipment.

5. PRECAUTIONS

8

3. Storage and maintenance

The equipment should be handled carefully and lightly. Be sure that it installed or kept in a cool, dry and ventilated place.
Don't store the machine together with the articles that are combustible, poisonous, caustic, and explosive.
This equipment should be stored in a room where the relative humidity is
≤80%, atmospheric pressure is 50kPa to 106kPa, and the temperature is -10°C +- 50°C.
If not use for a long time, please make the machine get through to the power and water once per month for five minutes.

4. Transportation

1. Excessive impact and shake must be prevented in transportation, lay it carefully and lightly and don't invert it.
2. Don't put it together with dangerous goods during transportation.
3. Avoid solarization and getting wet in rain or snow during transportation.

5. Working condition

fig 6
We offer one year warranty to the equipment according to the warranty card. The repair of the equipment should be carried out by
our professional technician. we are not responsible for any irretrievable damage caused by the non-professional person.
Environment temperature :
Relative humidity :
Atmosphere pressure :

6. AFTER SERVICE

7. SYMBOL INSTRUCTION

Class II equipment
Type BF applied part
Alternating current
Socket for the foot switch
Adjustment for the water flow
Temperature limitation
Humidity limitation
Date of manufacture
Authorised Representative in
the EUROPEAN COMMUNITY
9
GB
5°C to 40°C
≤80%
70 kPa to 106 kPa
Caution, consult accompanying
documents
Use indoor only
Can be autoclaved
30VDC power supply socket
Water entrance pressure 0,01
MPa - 0,5 MPa
FDA marked product
CE marked product
Atmospheric pressure
for storage
Manufacturer

Publicité

loading