Note
Electromagnetic compatibility (EMC)
You can use this device in industrial environments, households and small businesses.
For metal housings there is an increased electromagnetic compatibility compared to high-frequency radiation. This
protection can be increased by grounding the housing, see Chapter "Connecting (Page 31)".
Note
Improvement of interference immunity
● Lay signal cables separately to cables with voltages > 60 V.
● Use cable with twisted wires.
● Keep the device and the cables at a distance from strong electromagnetic fields.
5.2
Connecting the device
①
Atmospheric pressure
②
Vent pipe
Figure 5-1
Pressure transmitter, connection diagram
The pressure transmitter is connected to the cable box.
1. Connect the cable of the pressure transmitter to the terminals 1 (-), 2 (+) and the grounding.
2. The vent pipe must be connected to the atmosphere inside the box.
①
Brown
②
White
Figure 5-2
Pressure transmitter, electrical connection
32
③
Transmitter
④
Cable box
③
Green
④
Protective conductor connection/equipotential bonding
SITRANS LH100 (7MF1572..) Transmitter for hydrostatic level
A5E32588610-02, 02/2014