Table Des Matières; Contraindications; Warnings - DENTSPLY Cavitron BOBCAT Pro Mode D'emploi

Masquer les pouces Voir aussi pour Cavitron BOBCAT Pro:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8
Table of Contents
Section Section Title
Number
Description of Contents
1
Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2
Contraindications and Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1
2.2
3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.1
Precautions for All Systems
3.2
4
Infection Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4.1
Infection Control Information Reference Card
4.2
4.3
5
Installation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
5.1
5.2
5.3
5.4
5.5
5.6
Power Cord/Power Connection
5.7
Rear Panel Controls
5.8
BOBCAT ® Pro Ultrasonic Scaler Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
6
6.1
6.2
6.3
6.4
6.5
7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
8
Techniques for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
8.1
8.2
8.3
9
9.1
Daily Maintenance-Start-Up Procedures, Between Patients, Shut-Down Procedures
9.2
10
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
10.1
10.2
11
Warranty Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
12
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11
13
Equipment Classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
14
Disposal of Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
ENGLISH • 2
Section 1: Indications
Ultrasonic procedures:
• All general supra and subgingival scaling applications.
• Periodontal debridement for all types of periodontal diseases.
Page
• Endodontic procedures.
Number
Section 2: Contraindications and Warnings

2.1 Contraindications

• Ultrasonic Scalers should not be used for restorative
dental procedures involving the condensation of
amalgam.

2.2 Warnings

• Persons fitted with cardiac pacemakers, defibrillators
and other active implanted medical devices, have been
cautioned that some types of electronic equipment might
interfere with the operation of the device. Although no
instance of interference has ever been reported to
DENTSPLY, we recommend that the handpiece and
cables be kept at least 6 to 9 inches (15 to 23 cm) away
from any device and their leads during use.
• There are a variety of pacemakers and other medically
implanted devices on the market. Clinicians should
contact the device manufacturer or the patient's
physician for detailed information about the device.
• The use of High Volume Saliva Evacuation to reduce
the quantity of aerosols released during treatment is
highly recommended.
• It is the responsibility of the Dental Healthcare
Professional to determine the appropriate uses of this
product and to understand:
- the health of each patient,
- the dental procedures being undertaken,
- and applicable industry and governmental agency
recommendations for infection control in dental health
care settings,
Section 3: Precautions
3.1 Precautions for All Systems
• Do not place the ultrasonic scaler on or next to a
radiator or other heat source. Excessive heat may
damage the ultrasonic scaler's electronics. Place the
ultrasonic scaler where air is free to circulate on all sides
and beneath it. Do not cover vents on rear panel.
• The ultrasonic scaler is portable, but must be handled
with care when moving.
• Equipment flushing and dental water supply system
maintenance are strongly recommended. See Section 9:
System Care.
- requirements, and regulations for safe practice of
dentistry; and
- these Directions for Use in their entirety, including
Section 3 Precautions, Section 4 Infection Control, and
Section 9 System Care.
• Where asepsis is required or deemed appropriate in
the best professional judgment of the Dental Healthcare
Professional, this product should not be used.
• During boil-water advisories, this product should not be
operated as an open water system (e.g. connected to a pub-
lic water system). A Dental Healthcare Professional should
disconnect the system from the central water source. The
Cavitron DualSelect™ system can be attached to this
unit and operated as a closed system until the advisory is
cancelled. When the advisory is cancelled, flush all incom-
ing waterlines from the public water system (e.g. faucets,
waterlines and dental equipment) in accordance with the
manufacturer's instructions for a minimum of 5 minutes.
• Prior to beginning treatment, patients should rinse with
an antimicrobial such as Chlorhexidine Gluconate
0.12%. Rinsing with an antimicrobial reduces the chance
of infection and reduces the number of microorganisms
released in the form of aerosols during treatment.
• Failure to follow recommendations for environmental
operating conditions, including input water temperature,
could result in injury to patients or users.
• Close the water shut-off valve in the dental water
supply system every night before leaving the office.
• The use of an in-line water filter is recommended.
• Never operate the ultrasonic scaler without water
flowing through the handpiece.
• Grounding reliability can only be achieved when the
equipment is connected to an equivalent receptacle
marked "Hospital Only" or "Hospital Grade".
ENGLISH • 3