Setup And Operation; Device Setup; Operation - Sunoptic Surgical Titan X400 Mode D'emploi

Source lumineuse au xénon avec régulateur manuel d'intensité
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26

6. SETUP AND OPERATION

6.1 Device Setup

Place the Illuminator on a stable surface such as a cart, counter, stand, etc.
WARNING: DO NOT use the device in any environment with explosive or flammable gases.
WARNING: Avoid placing the device in an area where the illuminator may be splashed with liquids. A
clear space minimum of 5 inches (12.7 cm) behind and above the cabinet is required. The illuminator
should not be placed where its exhaust will influence other devices, nor where exhaust from other
devices will influence the illuminator.
WARNING: DO NOT obstruct the exhaust or cooling vents of the illuminator. User must ensure that
environment air temperatures surrounding the unit are within the allowable limits.
Make sure the power switch is in the OFF position.
Connect AC power cord to the appliance inlet located on the rear panel of the light source.
CAUTION: Use only power cords provided with the unit or cables approved for medical use.
WARNING: To prevent electric shock, connect power cords of peripheral equipment through medical
isolation transformers.
NOTE: When using a medical isolation transformer, ensure that the transformer has sufficient power
ratings. Ensure that the power cord is connected to mains power with a three-prong plug.
Connect a fiberoptic light guide to the topmost position of the rotating turret, ensuring that the
fiberoptic end tip type matches the accepting port.

6.2 Operation

After the power cable and light guide are properly connected, ensure the illuminator is in the lowest
dimming position, and turn on the illuminator by pressing the mains power switch located on the
front panel. The power indicator light within the switch should illuminate.
Ensuring that the light output from the attached light guide is directed to a safe area, adjust the
intensity control slider until the desired brightness is achieved.
NOTE: The brightness control slider operates a progressive mechanical shutter. The lamp is running
at full power irregardless of the intensity.
The hour meter located on the rear of the unit shows the total elapsed running time of the unit.
When the lamp is replaced, reset the hour meter to track total lamp usage hours on the replacement
lamp.
LIT-259 Sunoptic Surgical
Rev. B
®
Page 9 of 100
English)

Publicité

Table des Matières
loading

Table des Matières