3. SPECIFICATIONS

PARAMETER
Light Source Type
Color Temperature
Lamp Life
Lamp Replacement
Brightness Control
Light guide adaptor
Input Power
Circuit Protection
Operating Conditions
Storage Conditions
Dimensions
Weight

4. CLASSIFICATION

PARAMETER
System Classification
Isolation
EMC Certifications
CE Marking
Degree of protection against
harmful ingress of water
Degree of safety in the
presence of Flammable
Anesthetics
Mode of operation
LIT-259 Sunoptic Surgical
Rev. B
Ceramic 400W Xenon
5600 K (typ.)
1000 hours (typ.)
Cartridge replacement
Mechanical iris control
Rotating turret with STORZ, ACMI, WOLF, and OLYMPUS
100-240 VAC, 50/60 Hz
750W (max.)
Resettable
68 to 104⁰F (20 to 40⁰C), 30 to 85% RH non-condensing, 700 to
1060 hPa
-4 to 140⁰F (-20 to 60⁰C), 0 to 95% RH non-condensing, 700 to
1060 hPa
13.3" x 6.1" x 18" (W x H x D)
33.8 x 15.5 x 45.7 cm (W x H x D)
22 lbs / 10.0 kg
FDA Class I, 510(k) exempt
EU Class I, Active device per Annex IX, rule 1
Type BF
CISPR 11 Class A, IEC 60601-1-2 4
Electrostatic discharge: ±8 kV contact, ±15 kV air
Radiated RF EM Fields: 3 V/m, 80 – 2700 MHz
EFT / Burst: ±2 kV, ±1 kV signal lines, 100 kHz
SURGE: ±0.5, ±1 kV
Conducted disturbance: 3 V 150 kHz – 80 MHz and 6 V in ISM
bands
Power frequency magnetic fields: 30 A/m
Voltage dips: 0% Un / 0.5 cycles at 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°, 40 % Un / 5 cycles, 70 % Un / 25 cycles
Voltage interruptions: 0% for 5000 ms
Proximity fields: in accordance with EN 60601-1-2: 2015 table 9
Regulation (EU) 2017/745
IPX-0; no protection.
Equipment is NOT suitable for use in the presence of
flammable anesthetics.