6.2. Operation

Rotate the Dimming Knob located on the Battery Holster clockwise to turn on the device.
NOTE: If the Battery Holster beeps twice and fails to turn on, an incorrect battery pack has been
installed, and should be changed to the correct pack.
Continuing to rotate the Dimming Knob clockwise will increase the intensity of the LED,
while rotating counterclockwise will dim the LED.
Fully rotating the Dimming Knob counterclockwise until it clicks will turn off the device.
NOTE: Lower light intensities will result in longer runtimes.
The spot size of the headlight may be adjusted by rotating the Iris Ring located on the
headlight.
The Headlight Module position may be adjusted my manipulating the Linkages
connected to the Headband.
The Battery Holster will produce a single audible beep when the battery is running low
on charge. Two beeps signify that the battery is very low and should be replaced with
a fully charged battery.
CAUTION: To avoid over-discharging of the battery pack, install the battery pack in an approved
charger once the Battery Holster begins to notify of low charge. DO NOT store battery packs with
low charge for extended periods of time, and DO NOT leave discharged batteries in the Battery
Holster.
To disconnect the Headlight Module from the Battery Holster, press the retaining button
located on the power cord connectors and disengage.
When storing the device, reinstall the protective lens cap on the Headlight Module to
prevent damage to the Headlight Module lens.
6.3. Battery Pack and Charger Information
Please refer to the separate Battery Pack and Charger Instructions for Use on how to safely
handle and operate these devices.

7. CLEANING AND DISINFECTION

The headlight module and headband can be wiped down with commercially available cleansers
commonly used for disinfection of electronic equipment in hospitals such as ethyl or isopropyl
alcohols, disinfecting sprays containing quaternary ammonium compounds, or hydrogen peroxide.
The Headlight Module lens should be cleaned only with lens tissue.
WARNING: DO NOT use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid, or similar products. DO NOT use acetone, methyl ethyl ketone, or
halogenated / chlorinated hydrocarbon solvents or cleansers containing any of these restricted
compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids onto the device. Do
not allow liquids to enter the device seams or ventilation openings.
Follow all applicable bloodborne pathogen procedures as required by OSHA and/or your
hospital when cleaning and disinfecting the product.
WARNING: The Headlight is not sterilizable. DO NOT autoclave the Headlight Module,
Headband, or Battery Pack.
LIT-246
Sunoptic Surgical
Rev. E
®
Page 8 of 80
(English)