Connection
Connection Notes
Before opening the lid, ensure there are no dust deposits around the SITRANS LPS200.
Ensure that the atmosphere around SITRANS LPS200 is settled.
Use a fuse for the signal output (max. 6 A).
Make sure the main voltage does not exceed the maximum voltage listed on the
product label and the setting of the voltage selector.
In case of inexpert handling or handling malpractice, the electric safety of the
device cannot be guaranteed.
Provide protection for relay contacts to protect the device against spikes if inductive
loads are connected.
Switching Logic
7 6 5
Page EN-10
7
6
5
alarm output
relay
SITRANS LPS200 – INSTRUCTION MANUAL
M
1
2
N
L
115 V AC, 50 to 60 Hz, 4 VA
230 V AC, 50 Hz, 6 VA
or:
+
24 V DC, 2.5 W
All voltages ± 15 %
7 6 5
7ML19985FS61