Press the:
To set to the desired value.
Press on the:
The instrument then returns to the measurement mode.
NOTE: ADJUSTMENT RANGE: 0 to 8.
5. TABLE: CONDUCTIVITY OF POTASSIUM
5.1 CONDUCTIVITY VS. TEMPERATURE
TEMPERATURE KCl 0,1 mol.L
°C
15
16
17
18
19
20
21
22
23
24
25
Conductimeter C 320
CHLORIDE SOLUTIONS
Dilution
to 1/10
-1
KCl 0,02 mol.L
Conductivity
Conductivity
mS/cm
10.410
10.670
10.930
11.190
11.430
11.700
11.960
12.220
12.470
12.730
12.970
Dilution
Dilution
to 1/50
to 1/100
-1
KCl 0,01 mol.L
Conductivity
mS/cm
2.242
2.293
2.347
2.398
2.451
2.500
2.551
2.604
2.659
2.710
2.769
or
key to validate.
Dilution
to 1/1000
-1
KCl 0,001
mol.L
Conductivity
µS/cm
µS/cm
1 147
119.1
1 174
121.9
1 199
124.5
1 224
127.1
1 250
129.6
1 279
132.5
1 305
135.3
1 331
138.1
1 359
140.9
1 387
143.8
1 412
146.5
English
key.
-1
39