e. cleanInG your parI reusable nebulIzer
1. Disconnect the tubing from the air intake on
the bottom of the nebulizer cup.
2. if there is any moisture (condensation) in
the tubing, let the compressor run with only
the tubing attached for 2-3 minutes. the
warm air from the compressor will dry out
the tubing. alternatively, you can dry the
tubing by removing it from the compressor
and hanging it up with the "ends down" to
allow any moisture (condensation) to drain
out and evaporate. use a clean damp cloth
to wipe the exterior of the tubing.
3. Disassemble reusable nebulizer parts as
illustrated on page 3.
4. Wash all parts with warm water and liquid
dish soap. Do Not Wash the tubiNG.
5. Rinse the reusable nebulizer parts thoroughly
with warm water and shake out water.
D
6. air dry or hand dry nebulizer parts on a
clean, lint-free cloth. Reassemble nebulizer
when parts are dry and store.
7. You can also wash all parts of the reusable
nebulizer on the top shelf of a dishwasher.
Do Not Wash tubiNG. place the
reusable nebulizer parts in a dishwasher
basket, then place the basket on the top rack
of the dishwasher. Remove and dry the parts
when the dishwasher cycle is complete.
WARNING
to ReDuce RisK oF iNFectioN:
1. Nebulizer for single patient use (one person)
only. Do not share your nebulizer with other
patients.
2. You must regularly disinfect or sterilize the
nebulizer between treatments. Failure to do
so could lead to serious or fatal illness.
E
Questions? 1.800.Fast.neb (327.8632) | parI.com 6