K0320 Issue No. 1
To measure signals

Circuit continuity tester

To check wiring continuity use the
RTD input terminals identified with SW.
If NONE has been selected in the
output menu and SWITCH has been
selected in the input menu, press EXE
to enter the continuity check mode.
Closing the test loop will activate the
built-in buzzer top prove that the
wiring loop has a resistance of 2200
Ohms or less.
To calculate the switch resistance
press # to open the set-up menu and
select SWITCH RESISTANCE.
NOTE: Remote temperature
probe should not be connected in
this mode.