ervice, repair and parts

Recalibration of the MCX-II

Leave the cursor at NONE in the
112
INPUT and OUTPUT menu and press
the # key to open the settings menu.
Select CALIBRATION and press EXE.
Fill out the access code and press
EXE to open the calibration menu.
The access code is fixed at 9410
and works as an barrier to prevent
accidental access to MCX-II
calibration ranges.
The calibration menu shows all
relevant ranges to calibrate. Each
range can be recalibrated independent
of any other range. Only the cold
junction compensationsystem is
influenced by millivolts recalibration.
Select the ranges to be calibrated and
press EXE. Follow the instructions on
the next screens. Fill out the
calibration readings and press EXE to
confirm or to return to the calibration
menu.
Select an other range to continue
recalibration or press CE to leave
the calibration menu.
K0320 Issue No. 1