K0320 Issue No. 1
To measure signals
Ohms
Select OHMS and DIRECT in the
measure menu and press EXE to open
the working screen. The upper window
in the working screen in the indicates
the measured ohms reading.
The range is 0-2000 Ohms split into
two (auto)ranges of 0.01 to 400.00
Ohm and 400.0 to 2000.0 Ohm. The
exitation current is 0.9 mA.
The wiring can be 2, 3, or 4 wire.
The wire method is a true measurement
of voltage drop across unknown
resistance at a constant excitation
current.
Frequency
Select FREQ in the measure menu
and press EXE. Select FREQ and
DIRECT in the FREQ.menu and enter
a trigger level between 10 mV and 5V.
Press EXE to open the working
screen.
The upper window of the working
screen indicates the measured
frequency reading.
The range is 0-20.000 Hz split into
three (auto)ranges of 0-655 Hz,
655-1310 Hz and 1310-20.000 Hz.