D.1
Introduction
Cette section explique comment exécuter les tests qui démon-
trent au client que le système d'Orthoralix est conforme aux
normes indiquées.
D.2
Equipement requis
• Testeur d'adhérence
• Oscilloscope
• Multimètre
• Dosimètre
D.3
Tests de conformité aux normes
D.3.1
Etiquettes, indicateurs et
avertissements
1.
Vérifiez que l'étiquette indiquant la certification, l'identifica-
tion, le lieu et la date de fabrication sur le système d'éléva-
tion Z est apposée de façon permanente, lisible et aisément
accessible à la vue.
2.
Vérifiez que les facteurs techniques indiqués sur l'étiquette
du système d'élévation Z sur la colonne sont conformes aux
caractéristiques techniques.