7.
Installation of the contra-angle
Make sure the motor is stopped when installing the contra-angle.
Fig. 7.A
Connect the contra-angle to the handpiece properly. Use the original contra-angle only.
Fig. 7.B
The contra angle can be 340° rotated without take off, to make it easy to watch the screen during treatment.
Put the protective silicon cover onto the contra-angle. When removing after use, pull it straight out slowly.
• Check the proper connection by pulling the contra-angle gently, to avoid unexpected motor reverse
during use.
8.
Installation of the file
To install the file: insert the file and turn it, make sure it is correctly inserted.
Fig. 8.A
To remove the file: release the file while pressing the button.
• Make sure the motor is stopped before installing the file.
• Inspect the file head before inserting it. Do not use the file if its head is damaged.
• Pull the file gently to make sure it is securely connected.
• Insert or remove carefully the file to avoid finger injuries.
• File removal without pressing the button may damage the spindle.
9.
Dual Move control
9. 1. User interface (Fig. 9.A, 9.B, 9.C)
9.A.1
Contra-angle
9.A.2
Main switch (power on, motor start / stop)
9.A.3
Key for selection.
9.B
Standby screen
9.B.1
Battery level
9.B.2
Memory number
9.B.3
Running speed
9.B.4
Rotation direction
9. 2. Steps to operate
1. Power on / Power off with
To
power
Automatic
To power off: press simultaneously 9.A.2 (
2. Select memory program with < / > buttons (Fig. 9.D)
7 different customizable programs. M0: reciprocating mode. M1...M6: continuous rotation
* M0 Reciprocation mode is only available when specified on the box
4
(9.A.2 button)
on:
push
the
switch
off
if
Dual
• Never use reciprocating files in continuous rotation.
• Never use continuous rotary instruments in reciprocating mode.
9.A.4
< / - key to select a memory or other parameters
9.A.5
> / + key to select a memory or other parameters
9.A.6
User screen
9.B.5
Reverse Torque value
9.C
Display torque interface
9.C.1
Torque scale (N.cm)
9.C.2
Cursor of the reverse torque value
9.C.3
Real time torque display
9.A.2
button
(
),
Move
is
not
used
) and 9.A.4 (
) buttons.
the
standby
screen
for
a
while
(time-out
is
displayed.
to
be
set).