I. Technical description of the HEROfill
Oven
Technical description
®
The HEROfill
Oven is a medical device, class I, in compliance with MDD (Medical
Devices Directive). The oven must only be used in dental clinics of dental
®
personnel, and only for heating of HEROfill
Obturators. The product is classified
according to the DS EN 61010-1:2001, therefore, it must not be used within
reach of patients.
®
The HEROfill
Oven consists of an oven chamber heated by three ceramic
heating elements. An external power supply is attached to the oven.
®
The HEROfill
Oven has two heat settings: "High" (H) and "Low" (L). Use only
®
the "High" setting for traditional HEROfill
obturators.
Technical specifications
Power supply
Input power:
100–240VAC, 50/60Hz
Output power:
9VDC, 1.66A
Heating time
The heating time of the obturator varies with the temperature of the oven and is
between 1 and 3 minutes. Preheating takes 10-20 minutes depending of setting
Start using the oven
Connect the power supply to the oven's DC inlet and the
mains and switch on power.
A
Indicator
lights red while the oven preheats.
When the oven is ready, the indicator light turns green, and
you will hear a beep.
NOTICE: The Oven should remain connected to power all
day.
20