Description - Chauvin Arnoux C.A 5005 Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1

2 - DESCRIPTION

(See drawing in § 12 - APPENDIX)
1 TERMINALS
∅ 4 mm safety terminals
I COM: common, terminal that receives the black lead
I VΩ Ω Ω Ω Ω
: terminal that receives the red lead for voltages,
resistances and AC voltages with miniclamp MN 89
I I I I I 10 A DC: terminal that receives the red lead for the 10 A DC ranges
I I I I I mA DC: terminal that receives the red lead for the mA DC ranges
2
ANALOGUE DISPLAY
The dial comprises 6 scales:
I 2 black scales, with antiparallax mirror,
for V DC and AC (0.100 and 0.30) and for A DC (0.100)
I 1 green scale for Ω (0.10k)
I 2 red scales for A AC with miniclamp MN 89 (0.3 and 0.10)
I 1 red scale for dB (0.22)
3
"FUS" LIGHT
Neon for testing the 1 A and 10 A fuses on current measurement.
If the «Fus.» light is lit, change the faulty fuse(s).
NB: This light requires the presence of a voltage ≥ 80 V to come on.
4
"VOLTEST" LIGHT
Voltage presence light on ohmmeter.
If the «Voltest» light comes on: presence of a voltage
U > 15 V AC/DC. Unplug the leads and cut off the voltage
before making the resistance measurement.
5
SWITCH
Switch with 24 positions to select the functions and ranges.