5. Connector pins:
SH---shield wire
EXC+ --- Excitation +
EXC- --- Excitation -
SIG+ --- Signal +
SIG- --- Signal -
SEN+ --- Sense +, connect to EXC +
SEN- --- Sense -, connect to EXC -
Platform material:
4mm chequered plate
Wire the cable attached to the indicator as shown
1. Red Excitation+
1
2
1 RED
2. Black Excitation-
2 RED
3 RED
3
4
3. Green Signal-
4 RED
4. White Signal+
Error Messages
Error Message
0¯ ¯ ¯ ¯:
0_____:
_____:
¯ ¯ ¯ ¯:
CAL-Er:
Lo.bAt:
EEP.E0
EEP.E1
EEP.E2
1. Red Excitation +
2. Black Excitation -
3. Green Excitation -
4. White Signal +
Definition
Weight above range for calibrated
zero point.
Weight below range for calibrated
zero point.
Indicates an under range condition
Capacity exceeded
Calibration error
Low Battery
EEPROM can't be accessed
Configuration settings have
changed and not been stored
Configuration settings exceed
scale's normal range
Required Solution/Troubleshooting
Remove load before zeroing
Or
Recalibrate the scale.
Put platform on scale (if it has been
removed)
Or
Recalibrate the scale.
Recalibrate the scale.
Remove the load or a scale with a larger
capacity is required.
Restart calibration
Recharge the battery. Upon initial use, it is
recommended to charge battery for more
than 8 hours prior to use.
Replace S100 Indicator
Reconfigure and calibrate the scale to store
settings
Reconfigure and calibrate the scale to store
settings